beautypg.com

MTS Series 793 User Manual

Page 46

background image

MTS Series 793 Tuning and Calibration

About Calibration

Introduction

46

Extensometer calibration

Extensometers require special test fixtures to aid in calibration.

LVDT calibration

LVDTs are calibrated with a dial indicator or ruler. The indicator is mounted
between the actuator rod and a stationary point such as the load unit’s platen. A
typical LVDT has a positive and a negative output. This is usually considered as
tension and compression, but it is actually a motion each way from its
centerpoint. One output is calibrated with gain (typically compression) and the
other output is calibrated with delta K (typically tension).

Shunt calibration

Shunt calibration is a feature available for dc conditioners. It checks the integrity
of the conditioner/sensor combination. When a sensor range is calibrated, a shunt
calibration resistor is selected.

The system gives you the capability of verifying calibration accuracy by shunting
a precision resistor across one arm of the sensor’s Wheatstone bridge. The
resulting imbalance provides a reference value for later use.

m