beautypg.com

MTS Series 793 User Manual

Page 171

background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

171

2. Exercise the force standard.

Standalone FlexTest SE Controllers: Enable the front panel Dial, then
adjust the Setpoint to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis. Select Status >
Setpoint

Automated Controllers: Use the Manual Cmd slider on the Manual
Command
window to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis.

Example: When calibrating a ±10 kN range, exercise the actuator between 0
and -10 volts. To calibrate the same force sensor for a different range such as
±5 kN, exercise the load standard between 0 and -5 volts.

3. Apply a tensile force command that is 80% of the range’s full scale.

Standalone FlexTest SE Controllers:

A. Enable the front panel Dial.

B. Using the Dial, adjust the Setpoint for a tensile force command that is

80% of the full scale range.

C. Verify that your force transducer feedback signal is 80% of the full

scale range.

Automated Controllers:

A. Open the Manual Command window, and then click on the Enable

Manual Command.

B. Adjust the Manual Cmd slider on the Manual Command window for

a tensile force command that is 80% of the full scale range.

C. Use the Station Signals panel to verify that your force transducer

signal equals 80% of the full scale range.

During the initial calibration and tuning of your system, it may require
repeated adjustment for the tensile force command and feedback values to
match.