Task 6 establish the shunt calibration reference – MTS Series 793 User Manual
Page 210
MTS Series 793 Tuning and Calibration
Extensometer Calibration
Calibration Procedures
210
Task 6
Establish the shunt calibration reference
Each DC conditioner supports a shunt resistor. To establish the shunt reference
value, perform the following tasks.
1. Depending on your extensometer type, install a zero pin, gage block, or
special fixturing to mechanically hold the extensometer at its zero position.
2. Determine the shunt calibration resistor from the following table:
3. FlexTest SE/LE, FlexTest GT, TestStar IIm Controllers only: If you do not
have transducer ID modules on your sensor cables, install the shunt
calibration resistor as follows:
A. Select the appropriate shunt calibration resistor.
B. Bend the resistor leads 90º for a 0.3 inch separation.
C. Cut the resistor leads 0.12 inch from the bend.
D. Insert the resistor into the connector solder cups and solder.
E.
Complete and attach a shunt calibration label as specified on the
493.40/41 Carrier I/O Shunt Calibration Kit (MTS PN 100-028-185).
B
RIDGE
R
ESISTANCE
S
ENSITIVITY
R
ANGE
(% F
ULL
S
CALE
)
R
ESISTOR
V
ALUE
350
Ω
2 mV/V
100%
50%
20%
10%
49.9 k
100 k
249 k
499 k
350
Ω
1 mV/V
100%
50%
20%
10%
100 k
200 k
499 k
1000 k
700
Ω
2 mV/V
100%
50%
20%
10%
100 k
200 k
499 k
1000 k
700
Ω
1 mV/V
100%
50%
20%
10%
200 k
402 k
1000 k
2000 k