Task 6 establish the shunt calibration reference – MTS Series 793 User Manual

Page 210

background image

MTS Series 793 Tuning and Calibration

Extensometer Calibration

Calibration Procedures

210

Task 6

Establish the shunt calibration reference

Each DC conditioner supports a shunt resistor. To establish the shunt reference
value, perform the following tasks.

1. Depending on your extensometer type, install a zero pin, gage block, or

special fixturing to mechanically hold the extensometer at its zero position.

2. Determine the shunt calibration resistor from the following table:

3. FlexTest SE/LE, FlexTest GT, TestStar IIm Controllers only: If you do not

have transducer ID modules on your sensor cables, install the shunt
calibration resistor as follows:

A. Select the appropriate shunt calibration resistor.

B. Bend the resistor leads 90º for a 0.3 inch separation.

C. Cut the resistor leads 0.12 inch from the bend.

D. Insert the resistor into the connector solder cups and solder.

E.

Complete and attach a shunt calibration label as specified on the
493.40/41 Carrier I/O Shunt Calibration Kit (MTS PN 100-028-185).

B

RIDGE

R

ESISTANCE

S

ENSITIVITY

R

ANGE

(% F

ULL

S

CALE

)

R

ESISTOR

V

ALUE

350

Ω

2 mV/V

100%

50%
20%
10%

49.9 k

100 k
249 k
499 k

350

Ω

1 mV/V

100%

50%
20%
10%

100 k
200 k
499 k

1000 k

700

Ω

2 mV/V

100%

50%
20%
10%

100 k
200 k
499 k

1000 k

700

Ω

1 mV/V

100%

50%
20%
10%

200 k
402 k

1000 k
2000 k