Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

165

1. Select Gain/Delta-K for Cal Type.

Standalone FlexTest SE Controllers: Select Setup > Force > Calibration
> Cal Type > Gain/Delta-K

Automated Controllers: Select Gain/Delta-K for Cal Type on the
Calibration tab of the Inputs panel.

2. Exercise the force standard.

Standalone FlexTest SE Controllers: After enabling the front panel Dial,
adjust the Setpoint to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis. Select Status >
Setpoint

Automated Controllers: Use the Manual Cmd slider on the Manual
Command
window to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis.

Example: When calibrating a ±10 kN range, exercise the actuator between 0
and -10 volts. To calibrate the same force sensor for a different range such as
±5 kN, exercise the load standard between 0 and -5 volts.

3. Apply a tensile force command that is 80% of the full scale range.

Standalone FlexTest SE Controllers:

A. Using the Dial, adjust the Setpoint for a tensile force command that is

80% of the full scale range.

B. Verify that your force transducer feedback signal is 80% of the full

scale range.

Automated Controllers:

A. Open the Manual Command window, and then click on the Enable

Manual Command.

B. Adjust the Manual Cmd slider on the Manual Command window for

80% of the full scale range.

C. Use the Station Signals panel to verify that your force transducer

signal equals 80% of the full scale range.

During the initial calibration and tuning of your system, it may require
repeated adjustment for the tensile force command and feedback values to
match.