beautypg.com

MTS Series 793 User Manual

Page 183

background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

183

6. Determine the shunt calibration resistor from the following table:

Note

If you are calibrating a FlexTest IIm/CTC Controller, skip to step 9.

7. If you have sensor cables with optional transducer ID modules complete the

following procedure. If not, proceed to Step 8.

Install the shunt calibration resistor into the R9 location of the sensor ID
module. The sensor identification cartridge is molded into the sensor cable.

B

RIDGE

R

ESISTANCE

S

ENSITIVITY

R

ANGE

(% F

ULL

S

CALE

)

R

ESISTOR

V

ALUE

350

Ω

2 mV/V

100%
50%
20%
10%

49.9 k
100 k
249 k
499 k

350

Ω

1 mV/V

100%
50%
20%
10%

100 k
200 k
499 k
1000 k

700

Ω

2 mV/V

100%
50%
20%
10%

100 k
200 k
499 k
1000 k

700

Ω

1 mV/V

100%
50%
20%
10%

200 k
402 k
1000 k
2000 k

l

W1

W 2

1
2
3

SC

R6

R7

R8

R9

Solder the shunt cal
resistor here