beautypg.com

MTS Series 793 User Manual

Page 160

background image

MTS Series 793 Tuning and Calibration

Force Sensor Calibration

Calibration Procedures

160

4. Select the Calibration access level.

Standalone FlexTest SE Controllers: Select Config > Access Level >
Calibration

Automated Controllers: In the Station Manager window toolbar, select
Calibration

5. Set up a signal monitor.

See

page 134

for more information.

6. Mount the force standard.

7. If you are using Gain/Delta K or Gain/Linearization for calibration, set

initial conditioner calibration values on the Calibration submenu. A typical
complement of ranges could be: 100%, 50%, 20%, and 10% of full scale.
You can create ranges for any percentage of full scale. This creates a sensor
calibration file and sets up any ranges you may want.

Example: Suppose you have an actuator with a full-scale capacity of ±10
kN. You might create ranges for ±10 kN, ±5 kN, ±2.0 kN, and ±1.0 kN.

Note

Some systems do not provide or require multiple ranges, such as those
using full-range conditioners (e.g., Model 493.25 DUC module). In this
case, only one range is used (typically 100%).

Force Transducer

Load Standard

The illustration shows a load
standard in-line with the
force train coupled with the
force sensor.
The load standard reacts like
a stiff specimen. Be sure the
gain settings (PID) for the
control mode are
appropriate.
If the control mode has not
been tuned yet, use some
default values. Then
recalibrate the sensor after
the initial tuning.