beautypg.com

Task 4, “gain/delta k calibration, Task 4 gain/delta k calibration – MTS Series 793 User Manual

Page 201

background image

Extensometer Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

201

Task 4

Gain/Delta K Calibration

If you are using Gain/Delta K for your calibration type, complete the following
procedure. If not, complete Task 5 Gain/Linearization Calibration on page
204
.

Calibration setup

1. Ensure that the extensometer is properly attached to the calibrator frame

using the associated installation drawings and calibrator fixtures.

2. Set an initial transducer excitation voltage.

Standalone FlexTest SE Controllers:

Select Setup > Strain > Calibration > Cal Type > Gain/Delta K >
Excitation (peak)

Set the Excitation (peak) voltage value (typically 5-8 volts).

Automated Controllers:

A. In the Station Setup window’s navigation pane, locate and select the

Channels resource providing the strain feedback signal.

B. In the Station Setup window, click

.

C. In the Inputs panel, click the Calibration tab.

D. Set Type to Gain/Delta-K.

E.

Set Excitation (peak) voltage value (typically 5-8 volts).

3. If the extensometer uses a zero pin, remove the zero pin now.

4. Monitor the extensometer’s amplified output (strain).

Note

Before clicking Auto Zero, ensure that any offsets due to Auto Offset
and Manual Offset actions are eliminated.