beautypg.com

Task 6 gain/delta k calibration – MTS Series 793 User Manual

Page 142

background image

MTS Series 793 Tuning and Calibration

Displacement Sensor Calibration

Calibration Procedures

142

Task 6

Gain/Delta K Calibration

If you using Gain/Delta-K for your calibration type, complete the following
procedure. If not, complete Task 7 Gain/Linearization Calibration on page
147

Calibrate actuator

retraction

LVDTs can be calibrated so that a positive output represents actuator extension
and a negative output represents actuator retraction, or vice versa.

You calibrate the negative side of the output with gain and the positive side of the
output with Delta K.

Delta K compensates for differences in symmetry between the positive and
negative outputs.

You should calibrate actuator extension at 80% of the full scale range.

Gain controls

Pre-Amp gain is a selectable gain amplifier with predefined values. Since
changes in Pre-Amp gain can cause spikes in the feedback signal, Pre-Amp gain
can only be adjusted when hydraulics are off.

Post-Amp gain is a finer, operator-defined gain control that can be adjusted
when hydraulics are on.

mechanical input

Indicated output

Δ

K > 1

Δ

K = 1

Δ

K < 1

Gain