beautypg.com

MTS Series 793 User Manual

Page 181

background image

Force Sensor Calibration

MTS Series 793 Tuning and Calibration

Calibration Procedures

181

B. Enter the tension sensitivity value (+mV/V) in the Pos Tension entry

box.

C. Enter the compression sensitivity value (-mV/V) in the Neg

Compression entry box.

D. Enter the calibration excitation value (Vdc) in the Excitation entry

box.

Automated Controllers:

A. On the Inputs panel, enter the full scale force values in the Fullscale

Min/Max entry boxes.

B. Adjust Pos Tension for the required tension sensitivity value (+mV/V).

C. Adjust Neg Compression for the required compression sensitivity

value (-mV/V).

D. Adjust Excitation for the required calibration excitation value (Vdc).

mV/V Positive

Compression Calibration

Use the following procedure if your force transducer is set up so that a positive
output represents actuator extension (compression).

1. Select mV/V Pos Comp for your calibration type.

Standalone FlexTest SE Controllers: Select Setup > Force > Calibration
> Cal Type > mV/V Pos Comp

Automated Controllers: Select mV/V Pos Compression for Cal Type on
the Calibration tab of the Inputs panel

2. From the Calibration Data sheet for your force transducer enter the

following values on the Calibration submenu or tab:

Standalone FlexTest SE Controllers:

A. Enter the full scale force values in the Minimum and Maximum entry

boxes.

B. Enter the tension sensitivity value (-mV/V) in the Neg Tension entry

box.

C. Enter the compression sensitivity value (+mV/V) in the Pos

Compression entry box.

D. Enter the calibration excitation value (Vdc) in the Excitation entry

box.

Automated Controllers: