How to Perform Advanced Tuning Techniques

MTS Series 793 Tuning and Calibration

Tuning Procedures

103

Note

To improve your view of the waveform, click the Rescale button to
maximize the waveform on the display.

Using a square wave

Apply the tuning program with a square wave, and monitor the amplitude of the
settled portion of the error signal. The settled portion of the error signal should be
the same level for both segments, if not, adjust I Gain as required.

Using a ramp

Apply the tuning program with a ramp, and monitor the amplitude of the settled
portion of the error signal. The settled portion of the error signal should be
balanced, if not, adjust I Gain as required.

Upset recovery method

Monitor the error signal, then disrupt the system by changing the command by
adjusting the Setpoint control with the Dial. When you do this, the error should
return to zero within 5–10 seconds. If not, increase the reset (I Gain) setting and
repeat the procedure until the error zeros itself within a reasonable time period.

Adjust I Gain

Ideal Waveform

Balance difference
around zero

Zero