MTS Series 793 Tuning and Calibration
Shunt Calibration
Calibration Procedures
128
5. Select the control mode to which shunt calibration applies.
Standalone FlexTest SE Controllers:
Select Setup > force or strain> Calibration.
Automated Controllers:
A. In the Station Manager window’s Display menu, select Station
Setup.
B. In the Station Setup window’s navigation pane Channels, locate and
select the control mode needing a shunt calibration (e.g., force or
strain).
C. In the Station Setup window, click
.
6. Apply the shunt calibration.
Standalone FlexTest SE Controllers:
A. Locate the shunt calibration controls toward the bottom of the
calibration menu.
B. Apply the shunt by setting the Shunt State (+) control to On.
Automated Controllers:
A. In the Inputs panel, click the Shunt tab.
B. If desired, change units for the Shunt Reference Value.
C. To apply the shunt, click the button with three white bars.
While the shunt is applied, the bars will turn green
.
Note
To apply a shunt calibration to an auxiliary input, select Auxiliary Inputs
in the navigation pane and click the input signal needing shunt
calibration.
7. Compare the Current Shunt Value to the Shunt Reference Value.
Consider recalibrating the sensor/conditioner pair if the values differ by
more than 0.2%. For example, if the Shunt Reference Value is displayed in
Volts, consider recalibrating if the Current Shunt Value differs more than
20 mV.