beautypg.com

Task 8 save the calibration – MTS Series 793 User Manual

Page 154

background image

MTS Series 793 Tuning and Calibration

Displacement Sensor Calibration

Calibration Procedures

154

15. Verify linearization data.

Standalone FlexTest SE Controllers:

A. Using the front panel Dial, adjust the Setpoint for each retraction and

extension command point on the Linearization Data window.

B. At each command point, verify both the dial indicator value

(Standard) and its corresponding conditioner feedback value
(Conditioner).

C. Check validity before entering each pair of values on a new Calibration

Data Sheet.

Automated Controllers:

A. Adjust the Manual Cmd slider for each retraction and extension

command point on the Linearization Data window.

B. At each command point, verify both the dial indicator value

(Standard) and its corresponding conditioner feedback value
(Conditioner).

C. Check validity before entering each pair of values on a new Calibration

Data Sheet.

Task 8

Save the calibration

It is important that you save your sensor calibration values.

Standalone FlexTest SE Controllers: Select Setup > Open/Save Parameters >
<>

Automated Controllers: Click Save on the Calibration tab to save the current
calibration values to the sensor calibration file.

Note

Systems equipped with full range conditioners (e.g., Model 493.25 DUC
module) do not require or support multiple ranges. They use one range
that is typically set to 100%.