beautypg.com

Task 2 turn on hydraulic pressure – MTS Series 793 User Manual

Page 136

background image

MTS Series 793 Tuning and Calibration

Displacement Sensor Calibration

Calibration Procedures

136

Task 2

Turn on hydraulic pressure

This task activates the hydraulic pressure.

Do not place any part of your body in the path of a moving actuator.

A crush zone exists between the actuator and any equipment in the path of
its movement. Immediate and unexpected actuator response is possible
when you apply hydraulic pressure to your system.

Stay clear of the actuators when applying hydraulic pressure.

1. Ensure that displacement control mode is selected.

Standalone FlexTest SE Controllers: Select Status > Control Mode >
Disp.

Automated Controllers: Check Control Mode on the Station Manager
Function Generator panel.

2. Ensure that the Span is set for 100%.

Standalone FlexTest SE Controllers: Select Status > Span

Automated Controllers: Check the Master Span setting on the Station
Controls
panel

3. Check hydraulic and program interlock indicators.

Standalone FlexTest SE Controllers: If the Hydraulic Interlock or
Program Interlock indicator is lit on the front panel, determine the cause,
correct it, and then click Reset.

Automated Controllers: If the Interlock or Program indicator is lit on the
Station Controls panel, determine the cause, correct it, and then click Reset
or Reset/Override.

If either interlock indicator lights again, you will need to determine the
cause of the interlock and correct it before proceeding.

4. In the power selection box, click the Low button, and then High for the

HPU. If an HPU is not listed, start the HPU at the pump.

WARNING