3 glass vial in the dispenser rest position, 4 loading bath, 5 dispenser needles – Leica Biosystems CV5030 User Manual
Page 83: 6 cover slip catch tray, 7 cover slip magazine, 8 skids of the pick & place module, Cleaning and maintenance

83
Leica CV5030 – Robotic Coverslipper
7. Cleaning and Maintenance
7.5.3 Glass vial in the dispenser rest position
• Check the level, empty if necessary and add a compatible solvent.
7.5.4 Loading bath
• Fill the loading bath with a sufficient amount of solvent.
7.5.5 Dispenser needles
• Before starting up the instrument and before the priming cycle, check the dispenser needles
for proper flow and adhesive residue. Check the dispenser needles to ensure that they are
securely in place and not bent (
refer also to Chapter 4.6
).
7.5.6 Cover slip catch tray
• Remove glass residue from the cover slip catch tray. Ensure that the cover slip catch tray is
securely in place (
refer to Chapter 4.10
).
7.5 Description of the required daily cleaning measures (continued)
7.5.8 Skids of the Pick & Place module
• Check the skids on the bottom of the Pick & Place module for glued-on material (coverslip
mountant with glass shards and splinters) and, if necessary, carefully clean it with a lint-free
cloth soaked in compatible solvent.
7.5.7 Cover slip magazine
• Check that the coverslip magazine is correctly seated. Refilling the cover slips can result in
broken glass in or below the coverslip magazine, which must be removed.
Notice: If the cover slip catch tray is not correctly in place, collisions may result.
Notice: If the coverslip magazine is not correctly in place, there may be problems with
holding the cover slips during the coverslipping process.