Operation – Leica Biosystems CV5030 User Manual
Page 41

41
Leica CV5030 – Robotic Coverslipper
74
5.4 Brief inspection before starting the
coverslipping operation
65
68
Fig. 30
The following points have to be checked again
before starting the coverslipping operation:
• The level of the loading bath must be checked
before starting the coverslipping process. Fill
the loading bath with the corresponding sol-
vent. The level should reach up to the label
field of the specimen slide when a specimen
slide holder (74) is inserted.
• When inserting the loaded specimen slide
holder, ensure that the specimen slide re-
ceiving the cover slip has the side with the
specimen pointing towards the user (instru-
ment front).
5. Operation
• Push the loading drawer (68) with the loading
bath back into the instrument; close the load-
ing door (65) afterwards.
• Furthermore, check and, if necessary, correct
the following before starting the coverslipping
operation:
- Filling of the dispenser needle cleaner.
- Level of the coverslip mountant bottle.
- Sufficient filling and correct positioning of the
cover slip magazine.
- Presence of the output magazines.
- Selection of the desired parameter set (
refer to
Chapter 5.9
).