Coverslip Position Corr - correction of the cover slip placement position

5.11 Menu B – parameter settings

5. Operation

ð

Fig. 38

91

The cover slip placement position can be modi-

fied using the CSP (91) (cover slip placement

position) parameter (

see Fig. 38

).

If the cover slip placement position is to be

moved towards the edge of the specimen slide,

then select a negative value between –5 and –30.
If the cover slip placement position is to be

moved towards the label field, a positive value

between +5 and +30 has to be selected.
The changes can be made in increments of 5.

One increment corresponds to 0.5 mm.
The factory preset value is 0.

Stroke Position Corr - start position correction of the coverslip mountant stroke

The coverslip mountant stroke position can be

changed using the STP (starting point for the

coverslip mountant stroke) parameter.
If the starting point for the coverslip mountant

stroke is to be moved towards the edge of the

specimen slide, then select a negative value

between –10 and –100.
If the starting point for the coverslip mountant

stroke is to be moved towards the label field, a

positive value between +10 and +100 has to be

selected.
The changes can be made in increments of 10.

One increment corresponds to 1.0 mm.
The factory preset value is 0.

ð

Fig. 39

92