4.1 Installation site requirements (continued)
4.
Instrument Setup
• The instrument operator has to ensure that ESD safety precautions are maintained.
• The instrument operator is obligated to conform to the local workplace limit values and to document
them.
Furthermore, the instrument operator must ensure that there is sufficient air exchange and the
active carbon filter is changed at the recommended interval.
The instrument operator bears responsibility for complying with workplace limits and the measures
necessary for this, including documentation.
4.2 Unpacking the CV5030
Fig. A
1
2
3
4
7
Opening the package
(
Fig. A
)
• Unscrew the 8 screws (
Fig. A
,
2) on the sides
of the wooden box and loosen the cover.
• Carefully lift the cover from the wooden box.
When the instrument is delivered, check the tilt indicators on the pack-
aging. If the arrowhead is blue, the shipment was transported laying
flat, was tilted at too great an angle or fell over during transport.
Note this on the shipping documents and check the shipment for
possible damage.
Only personnel authorized by Leica may unpack and install the instrument.