Operation – Leica Biosystems CV5030 User Manual
Page 52

52
Instructions for Use V 3.0 RevE – 04/2013
5.7 Display indicators and instructions (continued)
5. Operation
Display
Explanation
Action
The instrument is in pause mode.
The coverslipping process was
interrupted due to the empty
cover slip magazine.
Fill the cover slip magazine.
Pressing START allows work to
continue.
The instrument is in pause mode.
There is no loading bath in the
loading drawer.
Insert a loading bath into the
loading drawer. Press RESPOND
as confirmation, then press
START to continue coverslip-
ping.
The instrument is in pause mode
since a vacuum could not be
attained in the system.
Check the coverslip mountant
bottle/dispenser system for
leaks and whether both O-rings
are present; replace the O-rings,
if necessary.
The instrument is in pause mode;
the working pressure in the sys-
tem could not be attained.
Check the coverslip mountant
bottle/dispenser system for
leaks and whether both O-rings
are present; replace the O-rings,
if necessary.
The instrument is in pause mode;
the system was unable to grab a
specimen slide.
In the specimen slide holder,
check whether the specimen
slides are inserted correctly
and undamaged. If the message
occurs repeatedly despite the
specimen slides being inserted
correctly, remove the specimen
slides and apply cover slips by
hand.