2 daily cleaning and maintenance - overview, Daily cleaning and maintenance - overview, Cleaning and maintenance – Leica Biosystems CV5030 User Manual
Page 80

80
Instructions for Use V 3.0 RevE – 04/2013
A
Check the loading chute and complete working area of the transport chain for broken glass
and adhesive residue. Carefully remove the adhesive residue and broken glass (
refer to
Chapter 7.5.1
).
B
Check the plastic container of the dispenser needle cleaner and, where applicable, refill
it with a maximum of 5 ml of a compatible solvent (
refer to Chapter 7.5.2
).
C
Check the level of the glass vial in the dispenser rest (park) position and, if necessary, fill
it with a maximum of 10 ml of a compatible solvent (
refer to Chapter 7.5.3
).
D
Fill the loading bath with a sufficient amount of solvent (
refer to Chapter 7.5.4
).
E
With the instrument switched on, prime the dispenser needle with coverslip mountant and
check the flow. If only a few drops or none at all come out of the dispenser needle, it is
blocked and must be exchanged for a new needle of the same type (
refer to Chapter 7.5.5
).
Insert the blocked dispenser needle into a solvent to release the blockage.
F
Inspect the cover slip catch tray (
refer to Chapter 7.5.6
) and the cover slip magazine (
refer
to Chapter 7.5.7
) for:
• Broken glass
• Correct seating
Refill the cover slip magazine if necessary.
Notes on cleaning agents
•
Clean the hoods and the housing with mild and pH-neutral commercial household cleaners.
The finished surfaces have only limited resistance to solvents!
• Clean the cover slip sensors, suction cups, skids of the Pick & Place module, transport
chain, specimen slide holder and output magazines using a lint-free cloth soaked in com-
patible solvent.
•
Never leave accessories lying in solvent or water for a long time (e.g. overnight) to prevent
damage.
7.1 Notes on cleaning and maintenance (continued)
7.2 Daily cleaning and maintenance - overview
7.
Cleaning and Maintenance