5.6 Interrupting the coverslipping operation
Press PAUSE to interrupt the coverslipping operation.
• The currently processed specimen slide is completely coverslipped
and pushed into the output magazine, accompanied by the message
"PAUSING".
• If this step is ended, the message "PAUSED" is shown in the display.
A signal tone is emitted at the same time.
• Confirm the signal tone by pressing the RESPOND button.
• Use the START button to continue the coverslipping operation.
ò
5. Operation
Whenever the instrument is to pause only briefly (e.g. to refill the
cover slip magazine), use PAUSE to interrupt the instrument with-
out having to run through a complete initialization and prime cycle.
If a just partly filled output magazine is to be removed, press LIFT
and proceed as outlined under Chap. 5.5.
In PAUSE mode, pressing and holding the RESPOND button
(>2 sec) allows you to change coverslipping parameters or select
another parameter set (SET 1 through 4) (
see Chap. 5.8
).