2 technical data, Technical data, Instrument components and specifications – Leica Biosystems CV5030 User Manual
Page 14

14
Instructions for Use V 3.0 RevE – 04/2013
3.
Instrument Components and Specifications
3.2 Technical data
Nominal supply voltages:
100 - 240 V AC, ± 10 %
Nominal frequency:
50 - 60 Hz
Approvals
CE, CSA C/US
Fuse:
Automatic fuseT5A
Nominal power:
100 VA
Dimensions (W x H x D):
Hood closed: 420 x 600 x 550 mm
Hood opened: 420 x 980 x 550 mm
Width (from the left to the right base):
370 mm
Depth (from the posterior base to the front base): 525 mm
Multistainer workstation:
Hood opened: 1620 x 980 x 760 mm
Weight (without packaging material):
approx. 57 kg
Net weight (including packaging material):
approx. 104 kg
Operating temperature range:
+15 °C - +35 °C
Storage temperature range:
–29 °C - +50 °C
Transport temperature range:
–29 °C - +50 °C
Relative humidity:
Maximum 85 % (non-condensing)
IEC 1010 classification:
Protection class 1
Pollution degree 2
Overvoltage installation category II
International protection class:
IP20
Operating elevation:
Up to a max. of 2000 m above sea level
A-weighted noise level:
<65 dB(A), measured in 1 m target range
Sockets
RS 232C, service interface
Uninterruptible power supply (UPS):
The uninterruptible power supply (UPS) should be
designed for a capacity of at least 200 VA over a
time frame of 5 minutes.
Heat dissipation:
100 J/s
Exhaust air:
Hose material
EVA (ethylene vinyl acetate)
Hose length:
3000 mm
Hose diameter:
32 mm
Hose circumference:
41 mm
Exhaust air quantity:
38.5 m
3
/h
Extraction:
Active carbon filter and extraction hose for con-
necting to an external extraction device