Operation – Leica Biosystems CV5030 User Manual
Page 53

53
Leica CV5030 – Robotic Coverslipper
5.7 Display indicators and instructions (continued)
5. Operation
Display
Explanation
Action
At least three cover slips in a
row are defective.
Check the cover slips in the
cover slip magazine.
If the messages occurs repeat-
edly despite the cover slips be-
ing intact, clean the cover slip
sensor. If the message persists,
clean the Pick & Place module
(skids and suction cups).
The user has pressed the STOP
button.
The output magazine, if loaded,
has to be removed from the
instrument. Grabbed specimen
slides or held cover slips also
must be removed from their
position by pressing the respec-
tive RELEASE SLIDE or RELEASE
COVERSLIP button.
Pressing START reinitializes the
instrument. Afterwards, proceed
the same as when switching
on the instrument (
refer to
Chapter 5.3.
).
Important!
If the instrument is operated as
a workstation, the user must en-
sure that before reinitialization,
no rack is in the transfer station.
The coverslipping operation
was ended. The number after
FINISHED corresponds to the
selected SET 1-4 program.
Remove the cover slipped speci-
men slides.