Instrument setup – Leica Biosystems CV5030 User Manual
Page 31

31
Leica CV5030 – Robotic Coverslipper
4. Instrument Setup
• After the dispenser needle cleaner is installed,
it must be aligned relative to the dispenser
needle (13).
• To do so, unscrew the screw (7) using the
No. 3 Allen key (6) (
Fig. 20/1
) and align the dis-
penser needle cleaner by moving the bracket
(3) sideways until the dispenser needles (13)
move into the correct position in the middle of
the cleaning port (
Fig. 21
).
• Finally, ensure that the gripper (14) does not
collide with the dispenser needle cleaner
(
Fig. 22
) during its movement or that of the
transported specimen slide (18).
• If the edge of the specimen slide touches the
lid of the brush holder, you can carry out a fine
adjustment after unscrewing the screws (9) on
the housing wall (
Fig. 19/2
). To do so, use the
available play in the bores.
13
14
18
Alignment to the dispenser needle
Fig. 22
Fig. 21
Checking for freedom of movement
7
3
Fill the container with 5 ml of solvent
using the provided plastic pipette
(
Fig. 17,
5). Regularly check the level of
the container. The solvent being used
must be compatible with the coverslip
mountant.
4.8 Dispenser needle cleaner (nozzle cleaner) (continued)