4 cleaning and maintenance as necessary, 2 dispenser needle cleaner (nozzle cleaner), Cleaning and maintenance as necessary – Leica Biosystems CV5030 User Manual
Page 82: Cleaning and maintenance

82
Instructions for Use V 3.0 RevE – 04/2013
7.
Cleaning and Maintenance
A
Filling the coverslipping mountant bottle:
• If possible, top up the coverslip mountant at the end of the working day so that
any air bubbles that arise during filling are degassed by the next working day.
• The coverslip mountant needs between 6-12 hours of resting time to degas (time
depends on the type of the coverslip mountant).
• If necessary, clean the bottle neck and both O-rings (blue = bottle neck and
black = dispenser group) to remove coverslip mountant residue.
B
Replace the active carbon filter if necessary, but no later than after three months (
refer
to Chapter 7.7.1
).
C
Check the transfer station for dirt and clean it if necessary:
• TS5015: Transfer arm (
refer to Chapter 7.7.3
) and carriage in y-direction.
• TS5025: Discharge chute (
refer to Chapter 7.7.2
) and transfer arm.
7.4 Cleaning and maintenance as necessary
7.5 Description of the required daily cleaning measures
7.5.1 Loading chute and bath transport with transport chain
• Check the loading chute and bath transport with transport chain for broken glass and adhe-
sive residue and carefully remove any dirt or debris.
• To clean adhesive residue, use a lint-free cloth soaked in compatible solvent. Loose glass
shards and splinters can be cleaned up using a commercially available vacuum cleaner.
7.5.2 Dispenser needle cleaner (nozzle cleaner)
• Fill the dispenser needle cleaner with approx. 5 ml of solvent.
• Check the brush. If the brush is hard or very dirty, replace it with a new brush.
There is a danger of cutting injuries during this cleaning step. Therefore, proceed with the
necessary caution.
Wear protective clothing!