Workstation operation – Leica Biosystems CV5030 User Manual
Page 76

76
Instructions for Use V 3.0 RevE – 04/2013
6.
Workstation operation
6.3 Important instructions for operation as workstation
Fig. 44
83
81
The deep bath, which can be identified by side
notches (see Fig.
64
), can be used in workstation
mode in combination with a silver/blue transfer
arm (
Fig. 45
) only. The transfer arm is in the
rear part of the TS5015 / TS5025 Transfer Sta-
tions. Before using the deep bath, please check
whether the silver/blue transfer arm is available.
Older instruments are equipped with a silver/sil-
ver transfer arm. The deep bath may not be used
in combination with this transfer arm!
• Specimen slide holders loaded man-
ually into the CV5030 must always be
removed immediately after process-
ing.
• The unloading chute (
Fig. 44
, 81) of
the transfer stations should be emp-
tied after the "TS FULL" message ap-
pears. At the latest, this should be
done after a signal tone that occurs
after the third specimen slide holder
and the "PAUSED" and "TS FULL"
messages.
• Ensure that the coverslip magazine
in the CV5030 is filled regularly.
This will prevent complete empty-
ing of the coverslip magazine, which
would cause the robotic coverslip-
per to stop working and go into
pause mode.
• Full output magazines must be re-
moved regularly, as otherwise con-
tinuous workstation operation is not
guaranteed.
Fig. 45