11 refilling consumables, Instrument setup – Leica Biosystems CV5030 User Manual
Page 34

34
Instructions for Use V 3.0 RevE – 04/2013
4.
Instrument Setup
• Fill the loading bath (
Fig. 27
, 68) with a solvent
compatible with the coverslip mountant. The
level should reach up to the specimen slide's
label field.
• Fill the glass vial (69) in the rest position with a
solvent compatible with the coverslip mountant
(approx. 10 ml). The dispenser needle must al-
ways be immersed in the solvent.
• The standard scope of delivery includes two
different cover slip magazines with different
widths (22/24 mm). Fill and insert the magazine
corresponding to the cover slip width being used
(possible cover slip lengths 40 - 60 mm) (
refer to
Chapter 9, "Optional Accessories"
).
Fig. 28
4.11 Refilling consumables
44
69
40
62
In order to be able to degas air
bubbles generated while filling the
mountant bottle (44), the coverslip
mountant must sit for 6 - 12 hours (the
period depends on the type of cover-
slip mountant).
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)