5.7 Display indicators and instructions
5. Operation
Display
Explanation
Action
Appears after switching on the
instrument. Indicates that the
instrument is being initialized.
Do not press any buttons; wait
until the device is fully initialized
and the next message appears.
The user is prompted to check
the loading bath.
Open the loading door, pull out
the loading drawer and loading
bath. Then check whether there
is a lid on the loading bath or
an empty rack inside it (remove
both if they are present).
The loading drawer has been
pulled out.
Push in the drawer and close the
loading door.
The user is prompted to move
the dispenser into the park
position.
Move the dispenser into the
prime position.
The user is prompted to press
PRIME.
Press the PRIME button.
Dispensing pressure is built up
in the coverslip mountant bottle.
Wait until dispensing pressure
has been reached in the cov-
erslip mountant bottle and the
pump noise is no longer audible.
The display of the Leica CV5030 is used to display status and error messages that provide in-
formation about the current status of the instrument. Status messages always appear in the
bottom line of the display. The top line displays the operating status of the instrument. The
following table compiles all status messages. In addition, it indicates to the user which ac-
tion to carry out. All error messages displayed on the control unit of the device are listed in
Chapter 8.1
.