Workstation operation – Leica Biosystems CV5030 User Manual
Page 77

77
Leica CV5030 – Robotic Coverslipper
6. Workstation operation
6.3 Important instructions for operation as workstation (continued):
Important instructions for operation of different specimen slide holders
• Apart from workstation operation, the robotic
coverslipper can be manually loaded with
specimen slide holders from other manufac-
turers, or with other Leica specimen slide
holders, with use of a compatible bath insert
only.
• After coverslipping is concluded, ensure that
the specimen slide holder is removed once
again. Likewise, the replaced bath insert is to
be removed before starting a new workstation
operation.
The default specimen slide hold-
ers intended for workstation op-
eration are Leica plastic holders
for 30 specimen slides (Order
No. 14 0475 33643).
Use of other Leica specimen slide
holders (e.g. Leica plastic holders
for 20 specimen slides) or speci-
men slide holders from other man-
ufacturers is not permissible in
workstation operation.
Specimen slide holders from other
manufacturers are only permis-
sible when they have been tested
on the instrument. These specimen
slide holders, or compatible bath
inserts, are listed in
Chapter 9 -
"Optional Accessories"
. Impermis-
sible specimen slide holders may
in some cases not be recognized
by the instrument and a collision
may result.