5.8 Button functions for programming
Fig. 32
Press and hold
for 2 s
Press 1 x
Fig. 33
71
72
5. Operation
Press and hold RESPOND for approx. two seconds to set
the programs and parameter sets the instrument uses to
operate.
The individual parameter sets are set using two sub-
menus:
Menu A (
Fig. 32
) and Menu B (
Fig. 33
)
Toggle between the two submenus by briefly pressing
RESPOND. The active submenu (71) is displayed in the
right half of the top display line.
There are four selectable programs (SET 1 through 4)
that are displayed on the upper left half of the display
(
Fig. 32,
72).
The bottom line in the display shows the parameters in
MENU A:
STK - Length of the coverslip mountant stroke
VOL - Quantity of the coverslip mountant
TYP - Pressure in the mountant bottle
LEN - Length correction for the coverslip mountant stroke
MENU B includes the following parameters:
CSP - Cover slip placement position
STP - Starting point for the coverslip mountant stroke
SPK - Volume of the signal tone
LIM - Cover slip limit setting