1 setting up the cv5030, 3 preparing and adjusting the instrument, Preparing and adjusting the instrument – Leica Biosystems CV5030 User Manual
Page 20: Instrument setup

20
Instructions for Use V 3.0 RevE – 04/2013
• Remove the transport anchors.
• Insert the filter and attach the exhaust hose.
• Level the instrument.
• Install the dispenser group.
• Align the dispenser needle with the specimen
slide outfeed.
• Install the dispenser needle cleaner.
• Connect the power supply.
4.
Instrument Setup
• Grab hold of bottom of the instrument from
the front and rear (with at least 2 people; the
instrument weighs approx. 57 kg) and place it
on a stable laboratory table.
• When doing so, ensure that the instrument is
standing on all four feet.
• Pull the plastic protective hood off of the in-
strument by pulling upwards and remove the
two adhesive strips (20).
• Remove the foam safeguard (21) from the
output station.
• Open the loading door (65) and remove the
foam cover for the loading bath.
• Ensure that the provided accessories are
complete in accordance with the order.
Fig. 3
4.3 Preparing and adjusting the instrument
20
21
• Use the following accessories:
-
Mountant bottle
-
Cover slip catch tray
-
Cover slip magazine
-
Output magazine
-
Loading bath
-
Dispenser needle cleaner
-
Glass vial for dispenser rest position
-
Refill consumables
65
To commission the instrument, carry out the following tasks described in the chapters below:
4.2.1 Setting up the CV5030