Operation – Leica Biosystems CV5030 User Manual
Page 68

68
Instructions for Use V 3.0 RevE – 04/2013
5. Operation
5.13.1 Procedure (continued):
If artifacts become visible within the monitoring
process (streaks, cloudiness, discoloration), the
involved reagents must be inspected for compat-
ibility and storage life.
The staining protocol should be adapted as
needed.
7. Inputting the first test parameters:
7 a) Effect of parameters on the application
quantity
• STK - Stroke length of the coverslip mountant
corresponding to the coverslip length used.
• VOL - Quantity of the coverslip mountant:
1 (min.) -> 9 (max.)
• TYP - Pressure in the mountant bottle:
1 (min.) -> 10 (max.)
7 b) Guideline for low-viscosity coverslip moun-
tants:
Dispenser needle: 20 or 21 (smallest diameter)
Menu A:
• STK: Adjust stroke length to the cover slip
length used
(e.g.
STK = 50 for coverslip size 22 x 50 mm and
24 x 50 mm).
• VOL: Start the first coverslip trials with VOL set
low (approx. 2-4)
• TYP: Start the first coverslip trials with TYP set
low (approx. 2-4)
The parameters VOL and TYP are incrementally
adjusted (increase or reduction) until a suitable
application quantity on the specimen slide is at-
tained. Further procedures are described in the
following
Point 8, "Fine-tuning"
.
7 c) Guideline for high-viscosity coverslip moun-
tants
Dispenser needle: 18 or 16 (largest diameter)
Menu A:
• STK: Adjust stroke length to the cover slip
length used
(e.g.
STK = 50 for coverslip size 22 x 50 mm and
24 x 50 mm).
• VOL: Start the first coverslip trials with VOL
(approx. 5-7)
• TYP: Start the first coverslip trials with the
setting TYP (approx. 5-7)
The parameters VOL and TYP are incrementally
adjusted (increase or reduction) until a suitable
application quantity on the specimen slide is at-
tained. Further procedures are described in the
following
Point 8, "Fine-tuning"
.