8.MULTI-HOLD FUNCTION

68

8. MULTI-HOLD FUNCTION

By this function, up to 16 types of hold, graph and comparison set values can be stored

and selected with external switching signals of CODE0 - CODE3.

Normally, if there is no entry for CODE0 - CODE3, the set value of CH0 is selected, but

when CODE0 - CODE3 are in the following conditions, the set value of each setting CH

is selected.

(0: open, 1: short)

CODE3

CODE2

CODE1

CODE0

SettingCH

0

0

0

0

SettingCH00

0

0

0

1

SettingCH01

0

0

1

0

SettingCH02

0

0

1

1

SettingCH03

0

1

0

0

SettingCH04

0

1

0

1

SettingCH05

0

1

1

0

SettingCH06

0

1

1

1

SettingCH07

1

0

0

0

SettingCH08

1

0

0

1

SettingCH09

1

0

1

0

SettingCH10

1

0

1

1

SettingCH11

1

1

0

0

SettingCH12

1

1

0

1

SettingCH13

1

1

1

0

SettingCH14

1

1

1

1

SettingCH15

I/O terminal block

B1

COM2

B2

CODE0

B3

CODE1

B4

CODE2

B5

CODE3

Caution

It takes 15msec at the maximum for the

changed CH to become effective. During

this period, which CH is used for

comparison is undefined.

Also, when the CH is switched, the hold and

graph functions are reset to wait for starting

with the after-switching CH conditions

irrespective of the previous operation.