3. near zero, 4. hi-lo limit comparison mode, Near zero – FUTEK IPM600 (D610) Digital Display User Manual
Page 55: Hi-lo limit comparison mode, P44)
![background image](/manuals/372300/55/background.png)
6.COMPARISON FUNCTIONS
44
6-3. Near Zero
By this function, it is detected that the indicated value is near zero.
Near-zero ON:| indicated value | <=near zero set value
Near-zero OFF:| indicated value | > near zero set value
Setting range: 00000 - 99999
How to set
6-4. HI-LO Limit Comparison Mode
Set the operating condition of HI-LO limit comparison. Select the condition from the
following.
Continuous
HI-LO limit comparison is performed continuously.
MD
HI-LO limit comparison is performed when the indicated value is
stable.Set the stability parameters by the stability operation.
NZ
HI-LO limit comparison is performed when the indicated value is
not near zero.
Set the near zero parameters by the near zero operation.
MD+NZ
HI-LO limit comparison is performed when the indicated value is
stable and not near zero.
How to set
Turning on/off the near zero function is closely related to the auto print
function and HI-LO limit comparison.
For details, see the sections on HI-LO Limit Comparison Mode page 44
and Auto Print page 37 .
Setting call
Page 2
Comparison setting
Setting call
Page 2
Comparison setting