CAUTIONS
MOUNTING
TENSION LOADING MUST BE
IN LINE WHEN COMPENSATING
LINKAGES ARE NOT USED
LOAD MUST BE FLAT
AND PARALLEL WHEN
COMPENSATING LINKAGES
ARE NOT USED
1) DO NOT PULL ON OR CARRY SENSOR BY CABLE.
2) ANY TAMPERING OR REMOVAL OF COVER, CABLE AND/
OR CONNECTOR WILL VOID WARRANTY.
3) ALWAYS HAVE SENSOR PLUGGED IN DURING
INSTALLATION TO MONITOR OUTPUT TO AVOID
PERMANENT ZERO SHIFT AND OVERLOAD.
THREAD THE FIXTURE TO
THE SENSOR. THREADING
THE SENSOR TO THE
FIXTURE CAN APPLY TORQUE
THAT MAY DAMAGE THE UNIT
SENSOR CAN BE USED SANDWICHED
WITH MORE THAN ONE UNIT BETWEEN
TWO PLATES
ROD END BEARINGS WITH CLEVISES OR ALIGNMENT
COUPLERS MAY BE USED TO MAKE UP FOR
MISALIGNMENT
SENSOR CAN BE
USED TO MONITOR
VOLUME IN A HOLDING
TANK APPLICATION
LOAD MUST BE IN LINE AND
CENTERED
CLEVIS
ROD END
BEARING
A LOAD BUTTON MAY BE USED
WHILE THE BOTTOM OF THE
SENSOR IS SECURED TO A BASE
PLATE
CYLINDRICAL S-BEAM
STANDARD S-BEAM
ALIGNMENT
COUPLERS
1
3
5
7
8
6
4
2
JUNIOR S-BEAM