4-2. Actual Load Calibration Procedure

Follow the steps below to perform actual load calibration.

Release the calibration protection.

Set the calibration value No.

(This step may be omitted for use with only one

selection.)

Enter the minimum value of digital increments.

(This step may be omitted if there is no change.)

Determine the decimal point place of the indicated value.

(This step may be omitted if there is no change.)

Enter the zero point of the strain-gage sensor with no load

applied to the sensor.

Enter the span (gain) point of the strain-gage sensor with a

load applied to the sensor.

The calibrated value can be offset in advance.

(This step may be omitted if there is no change.)

Turn on the calibration protection for preventing

misoperation.

Protection Release

Increment

Load Calibration

Calibration Protection

Decimal Place

Digital Offset

4-9.

4-7.

4-8.

4-4.

4-6.

4-9.

Zero Calibration

4-3.

Value Selection

4-10.

Calibration

Actual

Calibration