5.SETTING OF FUNCTIONS

32

Digital Zero by means of Keys

The digital zero may be performed by short circuiting the D/Z to the COM2 on the rear

panel signal I/O terminal block.

5-2. Digital Filter

The digital filter is a function for reducing drifts of the indicated value by means of a

moving average of data converted from analog to digital. The number of the moving

averages can be selected a range between 0 and 512.

With an increase in the number of filterings, the indicated value becomes more stable,

but the response to inputs becomes slower.

Number of settings:OFF, 2, 4, 8, 16, 32, 64, 128, 256, 512

How to set

The value of a digital filter is scrolled with a

button.

D/Z

OFF

ON

Keep on for 1.0 msec. or more.

Note

Digital zero is reseted in case of power failure. Please set digital zero

again.

Setting call

Page 1

Operation Setting

* See the section on About a

Setting Call page 8 .