7.HOLD FUNCTIONS

49

7-1-3. Auto Start Level

If you select the time specified period mode with trigger, or relative maximum, relative

minimum or inflection point hold in the hold mode setting, set the start level.

Setting range: -99999 - 99999

How to set

7-1-4. Hold Point Shift

In the “sample hold” and “inflection point hold”, the sampling data is held retroactively

by the numerical value set under hold point shift.

Setting range: -99999 - 99999

How to set

Setting call

Page 1

Hold

In each hold mode of relative maximum/relative minimum/inflection

point, if the auto start level is set at 99999, the hold operation starts

only with the period signal H/M without performing level detection.

The actual hold point

t

w

When the hold point shift is set at 8,
the data eight pieces before
the actual hold point is held.

Caution

This does not work for other holds.

The retroactive sampling data is held irrespective of the hold

period. If the set numerical value is too large, point(s) out of the

period may be held.

Setting call

Page 2