7.HOLD FUNCTIONS

62

7-5-6. Inflection Point Hold

Detection starts when the indicated value crosses the auto start level with the H/M signal

inputted. Detection is performed as long as the H/M signal is on.

The hold is released by turning on the T/H signal as a reset signal.

The H/E output signal is on between the instant when the hold starts and the instant when

the T/H signal is turned on.

If the auto start level is not needed as a starting condition of the detection period, set

99999.

(Example) Inflection Point Hold

t1 A delay time between the instant when the indicated value exceeds the auto start

level and the instant when the value to be held is detected
0.5ms (max.)

t2

A delay time between the instant when the H/M signal is inputted and the instant

when the value to be held is detected
1.0ms (max.)

t3

A minimum reset signal width required for releasing the hold

1.0ms (min.)

Indicated Value

t3

Sensor Input Value

Detection

t

HI-LO Limit
Judging Output

Hold Period

Decide

H/M

t2

t1

Auto Start Level

T/H

H/E

OFF

ON

OFF

ON

OFF

ON

OFF

ON

HOLD

PEAK

ON

Blink

ON

OFF

ON

OFF

OFF

OFF

HOLD

Buttun

OFF

ON

or

OFF

Comes on if a value is held with conditions met in the

detection period, and goes out if a value is not held.