STANDARD FEATURES:
Up to 60 Conversions per second
Scaleable to 5 digits. ±99,999
Worldwide input power: 85 to 264 VAC
and 90 to 370 VDC (
Green
or
Blue
Connector)
Isolated selectable excitation to power sensors.
5VDC@200 mA, *10 VDC@120 mA
Peak hold and remote auto tare
Plug-in screw terminal
Weight 9 oz [0.25 Kg]
*10 VDC@120 mA is optional
IPM500/IBT500 QUICK START
Menu Button
Is used to scroll through
setting options and to
save individual settings.
Peak Button
Is used to select the dis-
played menu item. Pressing
the
Peak function on a me
item will enable the user to
change the items setting.
Reset Button
Is used to reset all display func-
tions. Selecting the Reset function
can change the numerical value of
by the
Peak function.
Alarms Button
Allows the user to scroll
backwards through the
menu options.
ENVIRONMENT:
OPERATING TEMP. 32°F to 130°F [0°C to 55°C]
STORAGE TEMP.
-100°F TO 180°F [-73
[[ °C t
C o
tt 82°C]
C
C
RELATIVE HUMIDITY 95% AT 100°F [38°C]
Panel
Securing
Screws
SPEC.
SYSTEM
TORQUE
SENSOR
PRESSURE
SENSOR
LOAD
CELL
&
FORCE
SENSOR
Signal Out
(Analog or
RS232 or
RS485
Power
connector
connector
connector
Power Supply
Scope or DMM
Data Acquisition Card
Computer
VAC 85 to 264 (Std)
VDC 9-37 (Opt)
connector
Futek VCal Sensor
OPTIONS:
* Analog Output
* Alarm/Relay
* RS232 / USB & TEDS
* RS485
* DC option (9-37 VDC)(Black or
Brown
Connector
)
*
RS232 / USB Software Available
1.772
[45 ]
+0.6
-0.0
+0.024
-0.0
3.622 [92
]
+0.032
-0.0
+0.8
-0.0
Panel Cutout
LED
Alarm 2
LED
Alarm 1
3.78 [96]
4.02 [102]
1.89
[48]
0.197
[5]
J1
J2
J3
BACK
J4
J5
IPM500 Also Available In Bench
Top Version (IBT500)
Power Source
Tare Switch
1
AC HI (+DC HI)
2
AC LO (DC RET)
3
EARTH GROUND
4
DIGITAL INPUT B
5
DIGITAL INPUT A
6
DIGITAL GROUND
Black
White
Green
J1
Power/Input Connections
1
Alarm1 - N/O Contact
2
Alarm1 - N/C Contact
3
Alarm1 - Common
4
Alarm2 - N/O Contact
5
Alarm2 - N/C Contact
6
Alarm2 - Common
J2
Alarm Output Connection
Alarm
Indicator
Power
Supply
1
0-20mA Output
2 0-10VDC Output
3 Isolated Ground
J4
- Excitation
1
- Sense
2
- Signal
3
+ Signal
4
+ Excitation
5
+ Sense
6
J5*
Sensor Connections,
6
-Wire
Black
White
Gry/Ylw
Red
Blue
Orange
- Excitation
1
- Sense
2
- Signal
3
+ Signal
4
+ Excitation
5
+ Sense
6
J5*
Black
White
Gry/Ylw
Red
Sensor Connections,
4
-Wire
RS232
or
RS485
J3
Analog Output
J5 Note:
* Color code shown below is per Futek standard wiring.
* Check your sensor color code per Futek spec drawing sheet.
Follow your sensor color code if it differs from the code below.
Jumper
Jumper
TEDS (IEEE 1451.4):
* TEDS option available for Futek IPM500
display, IBT500 display, VCal
TM
System
and all sensors.
* Futek also upgrades your existing sensors
with TEDS option.
* Call 1-800-23-FUTEK for details
or USB)
USB
or
or
TEDS
*
*
*
* For Serial # 0-227233