beautypg.com

6. alarm mode, Alarm mode, P46) – FUTEK IPM600 (D610) Digital Display User Manual

Page 57: How to set

background image

6.COMPARISON FUNCTIONS

46

6-6. Alarm Mode

The comparison target of the HI-HI limit and LO-LO limit set values can be changed

from “hold value” to “real-time value”. By this function, whether or not the indicated

value becomes abnormal during hold can be monitored.

Unavailable : Comparison operation is performed with hold values.

Available

: Comparison operation is performed with real-time values.

How to set

Hold values

HI-HI Limit HH

HI Limit HI

Real-time values

HH

HI

OFF

ON

OFF

HH

HI

OFF

Alarm mode unavailable

Alarm mode available

When the comparison output mode is “mode 0: HI limit operation for

all of HI-HI limit/HI limit/LO limit/ LO-LO limit”, the alarm mode

operation can only be performed with the HI-HI limit.

In a like manner, when the mode is “mode 4: LO limit operation for all

of HI-HI limit/HI limit/LO limit/LO-LO limit”, the alarm mode operation

can only be performed with the low-low limit.

Setting call

Page 2

Comparison setting