7.HOLD FUNCTIONS

55

7-4. Hold Setting --- mean value ---

7-4-1. Mean Value Sampling Count

If the mean value sampling count is set at 2 or more in the mean value hold, the

representative value of the sampling values by the set count (mean value by the count) is

adopted as the sampling data used for mean value calculation.

The maximum mean value detection time with the setting “1” is 5 sec., but the mean

value detection time can be extended by this setting.

Maximum mean value detection time

= Mean value sampling count × 5 [sec.],

where the number of updates of the mean value will decrease.

Number of updates of the mean value

= 2000 / Mean value sampling count [times/sec.]

Setting range: 1 - 999Times

How to set

Setting call

Page 2

Hold

About the maximum mean value detection time

Although the detection period is specified by the H/M signal, etc.,

detection cannot be carried out exceeding the maximum mean value

detection time set according to the mean value sampling count. If the

maximum mean value detection time is exceeded, detection ends

automatically, when the mean value is held.