5.SETTING OF FUNCTIONS

35

5-6. Zero Track

Gradual changes in the zero point due to drifts etc., are automatically tracked for

correction.

Setting range

ZT (time): 0.1 - 9.9 sec.

ZT (width): 01 - 99 markings

How to set

- When displacement of the zero point is within the set count of tracking

and it continues more than the set time, it is automatically made zero by

zero tracking function.

- The track delay time and the track band can be selected in the range

between 0.1 and 9.9 sec and between 00 and 99, respectively. When

the set value is 02, two graduations correspond to the indicated value.

When the time is set at 0.0 sec and the band is at 00, the zero track

function does not work.

Boundary of zero track

Band=count × minimum value of digital changes (minimum graduation) × 2

Indicated Value

DELAY

+COUNT

-COUNT

BAND

0

DELAY

From the point when it returned within the range,
counting will be resumed.

CAUTION

Since the zero track function works from the calibrated zero point of the

indicated value, it does not work if the indicated value is beyond the

track band. In such a case, obtain the zero point again by zero

calibration.