7.HOLD FUNCTIONS

66

7-6-3. Time Specified Period Hold

(Peak, Valley, Peak-to-Peak and Mean Value)

Hold is detected during the predetermined time (hold detection time) from the point in

time when the H/M signal is turned on. The hold is released by turning on the T/H signal

as a reset signal.

The H/E output signal is on between the instant when the H/M signal is turned off and

the instant when the T/H signal is turned on.

(Example) Time Specified Period Peak Hold

t1

A delay time between the instant when the H/M signal is inputted and the instant

when the value to be held is detected
1.0ms (max.)

t2

A delay time between the instant when the hold detection time is expired and the

instant when the value to be held is determined
1.0ms (max.)

t3

A minimum reset signal width required for releasing the hold

1.0ms (min.)

Indicated Value

t3

Sensor Input Value

Detection

t

HOLD

HI-LO Limit
Judging Output

H/E

Hold Period

t2

t1

Decide

H/M

T/H

Time

OFF

ON

OFF

ON

OFF

ON

OFF

ON

OFF

ON

OFF

ON

HOLD

PEAK

ON

Blink

ON

OFF

ON

OFF

OFF

OFF