16.D/A CONVERTER(OPTION)

101

16. D/A CONVERTER(OPTION)

A D/A converter is provided for obtaining analog output synchronized with the indicated

value of the F371.

The analog output ranges are -10 - 10V output and 4 - 20mA constant-current output.

By using the D/A zero setting and D/A full scale setting functions, analog output can be

obtained between zero (0V, 4mA) and full scale (10V, 20mA) with respect to the

predetermined digital value.

The output circuit and the main unit circuit are isolated. The resolution is 1/10000 to -10

- 10V, and the conversion rate is 2000 times/sec. The maximum voltage output range

are about ± 11V, the maximum current output range are about 2.4 - 21.6 mA.

- Taking voltage output signals

Connect external equipment (2K

or more load resistance) to VOL and G of the F371.

Voltage/current output terminals:
Terminals to extract voltage or
current signals.
VOL and CUR indicate signals
and G indicates ground.
-10 - 10V and 4 - 20mA can be
obtained between VOL and G and
between CUR and G, respectively.

D/A
CONVERTER

VOL

G

External equipment

Load resistance

SIG

GND

Inside Outside

F371

2K

or more