REMOTE CONTROL

99

Command Syntax

The commands are partially compatible with IEEE488.2 (1992) and SCPI

(1994) standard. Commands are NON-case sensitive.

Example command

conf:volt:dc _1

1

2 3

1: Command Header

2: Single space

3: Parameter

Parameter example Boolean

Boolean logic: 0 or 1. Used for On (1) or
Off (0) command.

NR1

Integer: 0, 1, 2, 3.....

NR2

Decimal number: 0.0, 0.1, 0.2,....

NR3

Floating point number: 4.5e-1, 8.5e+1,...

min, max

The GDM-8200A series automatically

translates to Minimum (min) or Maximum
(max) value available.

Automatic
parameter range
selection

The GDM-8200A series automatically translates the

command parameter into the closest available value.

Example 1

conf:volt:dc_1 (Sets the measurement
item to DC Voltage and the range to 1V).

the GDM-8200A series selects the 1V
range.

Example 2

conf:volt:dc_2 (Sets the measurement

item to DC Voltage and the range to 2V).
There is no 2V range so the

GDM-8200A series selects the closest
range, 10V.

Message
Terminator (EOL)

Remote
Command

Marks the end of a command line. The

following messages are in accordance
with IEEE488.2 standard.

LF, CR or CR+LF The most common

ELO character is
CR+LF.

Return
Message

LF+CR

The fixed and only

option.

Message Separator ; (semicolon)

Command Separator