Command syntax – GW Instek GDM-8200A User Manual User Manual
Page 99
![background image](/manuals/367425/99/background.png)
REMOTE CONTROL
99
Command Syntax
The commands are partially compatible with IEEE488.2 (1992) and SCPI
(1994) standard. Commands are NON-case sensitive.
Example command
conf:volt:dc _1
1
2 3
1: Command Header
2: Single space
3: Parameter
Parameter example Boolean
Boolean logic: 0 or 1. Used for On (1) or
Off (0) command.
NR1
Integer: 0, 1, 2, 3.....
NR2
Decimal number: 0.0, 0.1, 0.2,....
NR3
Floating point number: 4.5e-1, 8.5e+1,...
min, max
The GDM-8200A series automatically
translates to Minimum (min) or Maximum
(max) value available.
Automatic
parameter range
selection
The GDM-8200A series automatically translates the
command parameter into the closest available value.
Example 1
conf:volt:dc_1 (Sets the measurement
item to DC Voltage and the range to 1V).
the GDM-8200A series selects the 1V
range.
Example 2
conf:volt:dc_2 (Sets the measurement
item to DC Voltage and the range to 2V).
There is no 2V range so the
GDM-8200A series selects the closest
range, 10V.
Message
Terminator (EOL)
Remote
Command
Marks the end of a command line. The
following messages are in accordance
with IEEE488.2 standard.
LF, CR or CR+LF The most common
ELO character is
CR+LF.
Return
Message
LF+CR
The fixed and only
option.
Message Separator ; (semicolon)
Command Separator