Secondary display: configure2 command, Configure2 command – GW Instek GDM-8200A User Manual User Manual
Page 107
![background image](/manuals/367425/107/background.png)
REMOTE CONTROL
107
rout:chan
Configure channel in advanced mode.
Parameter: Channel, Function, Range, Auto Range
Example: rout:chan 101, 1, 2, 0 (Channel 101, Function 1
(DCV), Range 2 (DCV 1V), Disable Auto Range)
rout:chan?
Return channel configurations in advanced mode.
Parameter: Channel, Function, Range, Auto Range
Example: 101, 1, 2, 0 (Channel 101, Function 1 (DCV), Range
2 (DCV 1V), Disable Auto Range)
rout:del
Set delay timer for scan.
Parameter: 0 to 9999 (ms)
rout:del?
Return delay timer setting for scan.
Parameter: 0 to 9999 (ms)
rout:coun
Set number of count for scan.
Parameter: 1 to 999
rout:coun?
Return number of count for scan.
Parameter: 1 to 999
rout:func
Enable scan related functions.
Parameter: 0 (scan off), 1 (monitor), 2 (step), 3 (scan), 4
(advance)
rout:func?
Return scan related function status.
Parameter: 0 (scan off), 1 (monitor), 2 (step), 3 (scan)
Secondary display: CONFigure2 command
conf2:volt:dc
Configure 2
nd
display to DC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:dc_1 (DC Voltage, 1V range)
conf2:volt:ac
Configure 2
nd
display to AC Voltage.
Parameter: NR2, min, max
Example: conf2:volt:ac_1 (AC Voltage, 1V range)
conf2:curr:dc
Configure 2
nd
display to DC Current.
Parameter: NR2, min, max
Example: conf2:curr:dc_10e-3 (DC Current, 10mA range)
conf2:curr:ac
Configure 2
nd
display to AC Current.
Parameter: NR2, min, max
Example: conf2:curr:ac_10e-3 (AC Current, 10mA range)
conf2:res
Configure 2
nd
display to 2W Resistance.
Parameter: NR2, min, max
Example: conf2:res_10e2 (2W Resistance, 1kΩ range)
conf2:fres
Configure 2
nd
display to 4W Resistance.
Parameter: NR2, min, max
Example: conf2:fres_10e2 (Resistance, 1kΩ range)