APPENDIX

119

100.000MΩ 120.000MΩ

199.999MΩ

3.0% + 8

M

100.00Ω

120.00Ω

199.99Ω

0.05% + 5

1.0000kΩ

1.2000kΩ

1.9999kΩ

0.05% + 3

10.000kΩ

12.000kΩ

19.999kΩ

0.05% + 3

100.00kΩ

120.00kΩ

199.99kΩ

0.05% + 3

1.0000MΩ 1.2000MΩ

1.9999MΩ

0.05% + 3

10.000MΩ 12.000MΩ

19.999MΩ

1.5% + 3

100.00MΩ 120.00MΩ

199.99MΩ

5.0% + 5

F

100.0Ω

120.0Ω

199.9Ω

0.05% + 2

1.000kΩ

1.200kΩ

1.999kΩ

0.05% + 2

10.00kΩ

12.00kΩ

19.99kΩ

0.05% + 2

100.0kΩ

120.0kΩ

199.9kΩ

0.05% + 2

1.000MΩ

1.200MΩ

1.999MΩ

0.05% + 2

10.00MΩ

12.00MΩ

19.99MΩ

1.5% + 2

100.0MΩ

120.0MΩ

199.9MΩ

5.0% + 2

Diode/Continuity

Note

Max. Input: 500V DC or 500V rms AC

Item

Range

Diode

Approx. 2V, 0.5mA

Continuity

1 ~ 1000Ω

Frequency

Note

Max. Input: 750V rms or 1000V peak

Frequency

Sensitivity

Accuracy (reading%+digits)

10Hz ~ 100kHz 0.1V

0.05% + 15

100kHz ~ 600kHz 1V

0.05% + 3

600kHz ~ 800kHz 2.5V

0.05% + 3

Temperature

Note

Sensor errors excluded from Temperature specifications

Type

Measurement Range

Thermo Couple

K

0 ~ +300°C

T

0 ~ +300°C

J

0 ~ +300°C

Resolution

0.01°C (0 ~ 300°C)