Sense command – GW Instek GDM-8200A User Manual User Manual
Page 101
REMOTE CONTROL
101
conf:per
Sets measurement to Period and specifies range.
conf:cont
Sets measurement to Continuity.
conf:diod
Sets measurement to Diode.
conf:temp
Sets measurement to Temperature.
conf:stat:func?
Returns function of 1
st
display.
Parameter: 1 (DCV), 2 (ACV), 3 (DCA-10A), 4 (ACA-10A), 5
(DCA-mA), 6 (ACA-mA), 7 (2WR), 8 (Freq), 9 (TempC), 10
(AC+DCA-10A), 11 (AC+DCV), 12 (AC+DCA-mA), 13 (Diode),
14 (Period), 15 (TempF), 16 (4WR), 17 (Cont.)
conf:stat:rang?
Returns range of 1
st
display. Parameter:
DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
ACV: 1 (100mV), 2 (1V), 3(10V), 4(100V), 5(750V)
AC+DCV: 1 (100mV), 2 (1V), 3(10V), 4 (100V), 5 (1000V)
DCmA, ACmA, ACmA+DCmA: 1(10mA), 2(100mA), 3(1A)
2WR, 4WR: 1(100Ω), 2(1kΩ), 3(10kΩ), 4(100kΩ), 5(1MΩ),
6(10MΩ), 7(100MΩ)
DCA, ACA, AC+DCA (10A range): 1 (one range)
Freq, TempC, TempF, Diode, Period, Cont.: 1 (one range)
conf:auto
Set 1
st
display to Auto range.
Parameter: 0 (disable auto range), 1 (enable auto range)
conf:auto?
Return 1
st
display Auto range status.
Parameter: 0 (disable auto range), 1 (enable auto range)
SENSe command
sens:det:rate
Sets detection rate.
Parameter: s (slow), m (medium), f (fast)
Example: sens:det:rate_s (set detection rate to Slow)
sens:det:rate?
Returns detection rate.
Parameter: Slow, Mid, Fast
sens:temp:tco:type
Sets thermocouple type.
Parameter: j (type J), k (type K), t (type T)
Example: sens:temp:tco:type_j (set thermocouple type to J)
sens:temp:tco:type? Returns thermocouple type.
Parameter: J (type J), K (type K), T (type T)
sens:temp:rjun:sim
Set temperature simulation value.
Parameter: NR2
Example: sens:temp:rjun:sim_23
sens:temp:rjun:sim? Returns temperature simulation value.