beautypg.com

System related command, Status reporting command, Rs-232c interface command – GW Instek GDM-8200A User Manual User Manual

Page 105: Ieee 488.2 common command

background image

REMOTE CONTROL

105

trac:cle

Clears buffer contents.

SYStem related command

syst:disp

Turns display On or Off.
Parameter: Boolean
Example: disp_1 (display On)

syst:disp?

Returns display status, On of Off.
Parameter: Boolean

syst:beep:stat

Select beep mode.
Parameter: 0 (Off), 1 (Pass), 2 (Fail)
Example: syst:beep:stat_1 (Beep when pass)

syst:beep:stat?

Returns beep mode status.
Parameter: No beep, Beep on Pass, Beep on Fail

syst:err?

Returns current system error, if there is any.

syst:vers?

Returns system version.
Parameter: 1.00 ~

*rst

Reset system.

*idn?

Returns company name, model No., and system version.
Example: GW, GDM8255A, 1.0

STAtus reporting command

stat:ques:enab

Enable bits in the Questionable Data register.

stat:ques:enab?

Returns Questionable Data register contents in decimal
number.

stat:ques:even?

Returns Questionable Data event register contents in decimal
number.

stat:pres

Clear Questionable Data enable register.

RS-232C interface command

syst:loc

Enables front panel control and disables remote control

syst:rem

Enables remote control and disables front panel control

IEEE 488.2 common command

*cls

Clears event status register (Output Queue, Operation Event
Status, Questionable Event Status, Standard Event Status)

*ese?

Returns ESER (Event Status Enable Register) contents.
Example: 130 means ESER=10000010