beautypg.com

Ac/dc/ac+dc voltage measurement, E24), Select voltage range – GW Instek GDM-8200A User Manual User Manual

Page 24

background image

GDM-8200A Series User Manual

24

AC/DC/AC+DC Voltage Measurement

Voltage type

AC

0 ~ 750V

DC

0 ~ 1000V

AC+DC

0 ~ 1000V

AC

2

+DC

2

*AC+DC=

(AC = true RMS)

1. Activate ACV/
DCV

Press the ACV (AC Voltage) key or

DCV (DC Voltage) key.

or

For AC+DC Voltage, press the ACV
key and the DCV key together.

2. ACV/DCV
mode display
appears

AC(DC) + V

Indicates AC, DC, AC+DC Voltage

AUTO

Indicates Automatic range selection

100mV

2nd display shows the Voltage range

3. Connect the
test lead and
measure

Connect the test lead between the V
and the COM port. The display

updates the reading.

V

COM

Note

When measuring in1000V (maximum) range immediately
followed by 100mV (minimum) range, an error might

occur due to extreme range switching. In such case, take
at least one minute in between as an interval.